
Injections training by
Stanley F. Malamed, D.D.S. Video part 1 - Introduction partl 2 - techniqs error
Modified STA-INTRALIGAMENTARY INJECTION
The
STA System provides Dynamic Pressure Sensing™
technology that is capable of identifying specific tissues during the dental
injection. The STA mode allows the
practitioner to accurately identify the periodontal ligament tissue. It also enables the clinician to maintain the
correct needle position within the periodontal ligament when performing the
newly described STA Intraligamentary Injection. Developed by Dr. Mark Hochman,
the STA Intraligamentary Injection represents a new concept in local dental
anesthesia techniques.
The STA System is the only anesthesia
system that provides clinicians with the 3 critical elements of information
when performing a STA Intraligamentary Injection:
1.
It
guides the clinician to position the needle tip to the periodontal ligament.
2.
It
provides ongoing feedback that the needle has not moved during the procedure.
3.
It
alerts the dentist if there is a blockage or any leaks in the system.
Performing
the STA INTRALIGAMENTARY INJECTION
1. Turn the STA drive unit to “On”. The system will default to the STA mode.
2.
Load and attach the STA
bonded handpiece with the 30 gauge ½ inch needle and the appropriate
anesthetic. The unit will automatically
purge the air from the system. Rest the
handpiece in the cap holder.
3.
While holding the STA
handpiece in a pen-like grasp, place the needle into the gingival sulcus of the
tooth to be anesthetized. Simultaneously, activate the ControlFlo™ rate by depressing the foot control. It is important to gently and slowly advance
the needle within the sulcus, as if it was a periodontal probe. It is highly recommended that the clinician
use a finger rest to control and stabilize the needle movements.
4. The
STA System provides a continuous audible and visual feedback to guide the
needle tip to the periodontal ligament.
As the foot control is depressed, the device says “sensing” The user will then hear the word “Cruise” at
which time the cruise control function can be operated by removing one’s foot
from the pedal.
5. In
the STA mode, the DPS technology provides real time pressure feedback via:
a.
The
visual Pressure Sensing Scale (Gauge) is comprised of a series of orange,
yellow and green LED lights. The orange
LED’s indicate minimal pressure, the yellow LED indicate mild pressure and the
green LED’s indicate moderate pressures that are indicative of the periodontal
ligament tissue.
b.
The
auditory Pressure Sensing Scale is composed of a series of triple ascending
tones “beep, beep, beep” Increasing
pressure is indicated by the triple ascending sequence. When the periodontal ligament is identified,
the user will hear the letters “PDL” spoken three times, followed by a series
of extended tones “beeeep, beeeep” indicating correct needle positioning.

Fig.
13
The
continuous DPST technology provides the user with important on-going information
that the needle has not moved from the optimal location during the entire
injection process. The DPST feedback
will also alert the operator to the proper hand pressure to the handpiece. Excessive pressure can result in the
“blockage” of flow of anesthetic solution.
This will be detected and result in an “over pressure” condition.
The
“Over-pressure” condition, i.e. when pressures exceed 450 PSI, the over
pressure alarm will be sounded. The user will need to release the foot control
or stop cruise control. The user will
then need to restart the injection moving the needle to a new location. The unit’s maximum generated pressure is 450
psi ±
10%. The maximum infusion pressure will
not exceed 500 PSI.
Dosage volume
The
STA system allows for a virtually unlimited amount of anesthetic to be
deposited. The operator should use
his/her own judgment as the anesthetic drug selection and the volumes used. The
following are suggestions and clinicians are encouraged to reference the
appropriate drug manufacturers, current dental literature and textbooks for
recommended dosages and drug recommendations.
a.
A
drug volume of 0.9 ml is recommended for single rooted teeth.
b.
A
drug volume of 1.8 ml is recommended for multi-rooted teeth.
c.
The
use of local anesthetics containing a vasoconstrictor concentration of 1:50,000
parts is not recommended for an Intraligamentary injection.
d.
When
using local anesthetics concentrated at 4% (Articaine Hydrochloride 4% and
Prilocaine Hydrochloride 4%), a drug volume of 0.4 ml is recommended for single
rooted teeth and a drug volume of 0.9 ml is recommended for multi-rooted teeth.
Removal of the needle from the ligament
should be performed mid-way during the aspiration cycle. Since the injection is performed under
pressure, if the needle is otherwise removed, the patient’s mouth will be
sprayed with bitter tasting anesthesia.
Therefore, the operator is advised to remove the needle during
aspiration, i.e. when the STA drive unit is retracting during aspiration.
Based on the auditory and visual
pressure sensing feed-back, it is not unusual for the operator to have to
reposition the needle several times before locating the proper position of the
needle within the periodontal ligament. Additionally, slight needle movements can result in rapid
loss of pressure. The user will need to
withdraw and reposition the needle to establish an effective periodontal
ligament location.